Part Number : MODEL-1020 | Alternate P/N : MODEL1020 |
NSN : 6630-01-242-40506630012424050 | Item Name : Analyzer Sodium Potassium |
Manufacturer : Thermo Orion Inc | CAGE Code : 30260 |
NIIN : 012424050 |
MRC | Criteria | Characteristic |
---|---|---|
CYFZ | TEST SAMPLE/SPECIMEN SIZE | 100 MICROLITERS OF BLOOD OR 400 MICROLITERS OF PREDILUTED URINE |
AKYD | ACCESSORY COMPONENTS AND QUANTITY | POTASSIUM ELECTRODE,1; SODIUM ELECTRODE,1; REFERENCE ELECTRODE,1; BOTTLE OF STANDARD A,2; BOTTLE OF STANDARD B,1; WASTE BOTTLE,1; BOTTLE OF REFERENCE FILLING SOLUTION,1; BOTTLE OF TEST DYE SOLUTION,1; REFERENCE MEMBRANE ASSEMBLY,2; FLUID PATH MODULE,1; BOTTLE ELECTRODE RINSE SOLUTION,1; SLIDE VALVE,1; SAMPLE PROBE ASSEMLBY,1 |
CYFX | ELECTRODE TYPE | POTASSIUM/K+ AND REFERENCE AND SODIUM/NA+ |
CYFV | METHOD OF ANALYSIS | POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
ABRN | HEIGHT | 360.0 MILLIMETERS |
DPTH | DEPTH | 230.0 MILLIMETERS |
ELEC | VOLTAGE IN VOLTS | 120.0 OR 230.0 |
CYFY | ANALYSIS TEST/MEASUREMENT RANGE | SODIUM 100-170 MMOL/L IN BLOOD AND 14-1400 MMOL/L IN UNDILUTED URINE; POTASSIUM 2-7 MMOL/L IN BLOOD AND 0.4-400 MMOL/L IN UNDILUTED URINE |
CBBL | FEATURES PROVIDED | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
FEAT | SPECIAL FEATURES | IF THE ANALYZER DOES NOT CONTAIN AN INTEGRAL PRINTER,IT SHALL BE CAPABLE OF INTERFACING W/A COMMERCIALLY AVAILABLE EXTERNAL PRINTER |
WDTH | WIDTH | 260.0 MILLIMETERS |
CXCY | III PART NAME ASSIGNED BY CONTROLLING AGENCY | ANALYZER POTASSIUM AND SODIUM ION SELECTIVE: |
ACZB | FREQUENCY RATING | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
CYGE | TEST SAMPLE/SPECIMEN DELIVERY DEVICE | NEEDLE/PROBE, ANALYZER ASPIRATION SYSTEM |
CYHM | TEST SAMPLE/SPECIMEN TYPE | WHOLE BLOOD,SERUM,PLASMA,AND URINE |
ACDC | CURRENT TYPE | AC |