Part Number : N0-059 | Alternate P/N : N0059 |
NSN : 6630-01-364-85556630013648555 | Item Name : Analyzer Blood Gas |
Manufacturer : Mallinckrodt Sensor Systems Inc | CAGE Code : 0CNN1 |
NIIN : 013648555 |
MRC | Criteria | Characteristic |
---|---|---|
DPTH | Depth | 18.300 INCHES |
ELEC | Voltage In Volts | 120.0 AND 230.0 |
MRC | Decoded Requirement | Clear Text Reply |
CBBL | Features Provided | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
CXCY | Part Name Assigned By Controlling Agency | BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
ACDC | Current Type | AC |
CYFV | Method Of Analysis | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
ACZB | Frequency Rating | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
CYFY | Analysis Test/Measurement Range | PO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT. |
CYHM | Test Sample/Specimen Type | WHOLE BLOOD |
ABRN | Height | 8.400 INCHES |
WDTH | Width | 9.100 INCHES |
AQXY | Test Type For Which Designed | BLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT |
CYFZ | Test Sample/Specimen Size | 0.5 ML |
FEAT | Special Features | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |