Part Number : 31TA2655-2 | Alternate P/N : 31TA26552 |
NSN : 4920-00-128-85844920001288584 | Item Name : Test Stand Demand O |
Manufacturer : Carleton Life Support Technologies | CAGE Code : 99251 |
NIIN : 001288584 |
MRC | Criteria | Characteristic |
---|---|---|
SUPP | Supplementary Features | POWER 28VDC, 4 AMPS, SOURCE/BATTERY |
RDAL | Reference Data And Literature | T.O. 33D2-10-55-4; SERD NO. 0755 |
CXCY | Part Name Assigned By Controlling Agency | OXYGEN REGULATOR FIELD TESTER |
AGAV | End Item Identification | F-15 ACFT |
AAXX | Mounting Type | PORTABLE |
MRC | Decoded Requirement | Clear Text Reply |
FEAT | Special Features | 16.31 X 8.41 X 9.18; 24 LBS |
FTLD | Functional Description | USED TO VERIFY PROPER OPERATION OF THE OXYGEN REGULATOR WHICH C/O OF 9 TESTS: MAXIMUM FLOW, AIR CHECK VALVE OPERATION, SUCTION FLOW,AIR RATIO SHUTOFF, PRESSURE BREATHING, OUTLET LEAKAGE, DEMAND VAVE LEAKAGE, BLINKER FLOW AND MASK PRESSURE |